The simulation of linear accelerator Using BEAMnrc with DOSXYZnrc1
Ying-hua Shi,Ling-hong Zhou2,Xin Zhen,Wen-ting Lu Shu-xu Zhang School of Biomedical Engineering, Southern Medical Radiotherapy Center, Affiliated Tumor Hospital Guangzhou University Medical College Guangzhou, China Guangzhou, China Email: shiyh1029@
Abstract—Linear accelerator accurate simulation is the pleted. One doesn’t know the characteristics of the prerequisite of Monte Carlo dose calculation algorithm. The incident electron beams due to proprietary. And even the study simulated 6MV photon beam produced by Varian 600C details of same type of machines with specified photon energy medical linear accelerator to derive the mean energy and the full- are not the same due to positioning, machining and urate width half-maximum(FWHM) of incident electron beam intensity measuring of a linear accelerator. The FWHM of the electron distribution (assumed Gaussian distribution) with BEAMnrc and beam energy distribution even up to 20%, there is no DOSXYZnrc by matching calculated and measured depth dose correlation observed between calculated and measured depth curves and off-axis ratios at a depth of in water. Depth dose curves and off-axis ratios [11]. The influence of these dose curve is sensitive to the mean energy, insensitive to radial factors, only incide